ORIGINAL ARTICLE
Year : 2017  |  Volume : 23  |  Issue : 1  |  Page : 38-42

Corneal hysteresis values in normal Egyptian population


Department of Ophthalmology, Memorial Institute of Ophthalmology, Giza, Egypt

Date of Submission18-Sep-2015
Date of Acceptance17-Mar-2016
Date of Web Publication29-May-2017

Correspondence Address:
Alaa A Ali
13 Mohamed Ezz El-Arab Street, Cairo 11647
Egypt
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/1687-4625.207196

Rights and Permissions
  Abstract 

Introduction
Knowledge of the cornea's biomechanical properties is important for a wide variety of applications within ophthalmology. Corneal biomechanical variations are known to affect the accuracy of intraocular pressure measurements and may be used to identify early corneal disease and may assist in predicting refractive outcomes following corneal refractive surgery. It has also been suggested that corneal biomechanical properties may reflect globe biomechanics, and thus give an indication of the susceptibility of developing glaucomatous damage.
Aim of the work
The purpose of this study was to explore the relationships between Ocular Response Analyzer-generated corneal biomechanical characteristics and age in a sample of Egyptian population.
Materials and methods
Corneal hysteresis (CH) and corneal resistance factor (CRF) were measured in 195 Egyptians of different age groups.
Results
The mean CH value was 10.25 ± 0.12 mmHg (range, 6.5–14.4), and the mean CRF was 10.25 ± 0.15 mmHg (range, 4.9–14.2). The CH value was lower in older eyes, and the difference between the youngest age group (19–40 years) and the oldest age group (40–71 years) was statistically significant (t-test = 0.01). The mean CH in the youngest age group was 11.1 ± 0.14, and that in the oldest age group was 9.8 ± 0.21. The mean CRF in the youngest age group was 10.9 ± 0.18, and that in the oldest age group was 10.1 ± 0.19.
Conclusion
This study of corneal biomechanics in normal Egyptian eyes describes the interactions between age and Ocular Response Analyzer metrics. The study reveals that age is significantly associated with CH and CRF. This may help in diagnosis and treatment.

Keywords: corneal hysteresis, Egyptian, normal


How to cite this article:
Ali AA. Corneal hysteresis values in normal Egyptian population. Kasr Al Ainy Med J 2017;23:38-42

How to cite this URL:
Ali AA. Corneal hysteresis values in normal Egyptian population. Kasr Al Ainy Med J [serial online] 2017 [cited 2024 Mar 29];23:38-42. Available from: http://www.kamj.eg.net/text.asp?2017/23/1/38/207196


  Introduction Top


Ocular biomechanics is an increasingly important field. Overt corneal biomechanical problems have long been seen in keratoconus and corneal ectasia after corneal refractive surgery. In keratoconus, there are clear changes in the corneal collagen, and the cornea loses rigidity over time and becomes ectatic; in corneal ectasia, the ablation of some corneal stroma can weaken the cornea and result in progressive corneal deformation [1]. In refractive surgical practice, patients with pre-existing ectasia usually are excluded from treatment. However, individual variations in biomechanical integrity and postoperative wound healing preclude preoperative identification of all potentially vulnerable patients. There is considerable but mostly indirect evidence suggesting that the biomechanical corneal properties vary with age [2]. Quantifying the biomechanical corneal properties is difficult, but the available evidence supports corneal stiffening with age; in other words, there is an increment in Young's modulus [3], the ocular rigidity coefficient, that expresses the elastic properties of the globe [4], the cohesive tensile strength, and the breaking force of a tissue [5]. Young's modulus, also known as the tensile modulus, is a measure of the stiffness of an elastic material and is a parameter used to characterize elastic materials. Perhaps the single best descriptor of a given material's biomechanical properties at low strain is its Young's modulus (E), which is defined as the ratio of stress to strain, or where stress is an applied force (load/unit area), and strain is the deformation of the material to which stress has been applied (displacement/unit length): Young's modulus (E) = stress/strain. This parameter depends on the material's physical properties and dimensions. Importantly, when stress is applied and removed, elastic materials follow the same path during deformation and relaxation and ultimately recover the original shape. Viscoelastic materials, such as the cornea, can also recover the original shape after stress is removed, but the relaxation path differs from the deformation path; therefore, the relationship between stress and strain is nonlinear, and stiffening occurs as strain increases [6],[7] [Figure 1]. This behavior, referred to as corneal hysteresis (CH), results from dissipation of energy as heat in the material.
Figure 1: The relationship between stress and strain is linear in an elastic behavior and nonlinear in a viscoelastic behavior.

Click here to view


In fact, the effect of the corneal thickness on Goldmann applanation tonometer measurements may be less important compared with the effect of variations in corneal elasticity [4]. CH is a measure of the viscoelastic properties of the corneal tissue together with the corneal resistance factor (CRF) – that is, the 'energy absorption capability' of the cornea – and indicates the biomechanical integrity [8].

Knowledge of the cornea's biomechanical properties is important for a wide variety of applications within ophthalmology. Corneal biomechanical variations are known to affect the accuracy of intraocular pressure (IOP) measurements [9],[10] and may be used to identify early corneal disease [11],[12] and may assist in predicting refractive outcomes following corneal refractive surgery [13]. It has also been suggested that corneal biomechanical properties may reflect globe biomechanics, and thus give an indication of the susceptibility of developing glaucomatous damage [5].

Until recently, most investigations evaluating corneal biomechanics were based on ex-vivo tissue. However, the Reichert Ocular Response Analyzer (ORA; Reichert Ophthalmic Instruments, Buffalo, New York, USA) has facilitated an in-vivo measurement of aspects of corneal biomechanical properties. The device measures the central corneal response to indentation using a rapid jet of air and provides two metrics of corneal biomechanics, CH and the CRF. It is thought that CH predominantly reflects the viscous dampening properties of the cornea, whereas CRF, a metric empirically derived to be strongly correlated with central corneal thickness, is thought to be most associated with the cornea's elastic response [14]. However, how these metrics relate to conventional biomechanical measurements is still relatively unclear. Studies examining the ORA biomechanical metrics have suggested associations with age, central corneal thickness, and IOP [15],[16]; however, some suggest no association with these parameters.

The purpose of this study was to evaluate the relationships between ORA-generated corneal biomechanical characteristics and age in normal Egyptian eyes.


  Aim of the Work Top


The purpose of this study was to explore the relationships between ORA-generated corneal biomechanical characteristics and age in a sample of Egyptian population.


  Materials and Methods Top


Inclusion/exclusion criteria

Study participants comprised normal people attending the Memorial Institute of Ophthalmic Research for check-up. Data were collected between 2008 and 2012. Participants underwent a complete ophthalmic check-up to exclude any pathology. Participants were excluded from the study if they had any signs of corneal pathology, corneal astigmatism of 3 diopters or greater or a history of incisional or intraocular surgery, suspicious optic disc appearance, IOP of 20 mmHg or greater, a history of diabetes, or a family history of glaucoma in a first-degree relative. Soft-contact lens wearers were required to remove their lenses at least 24 h before study participation; rigid contact lens wearers were excluded from the study. For eligible participants, only one randomly chosen eye was measured for the study.

Procedure and data collection

Before instillation of topical anesthesia, participants underwent ORA measurements. Subsequently, three good quality waveform scans, defined as having symmetry in height between the two peaks of the waveform, were recorded, and the mean value was used in subsequent analysis. Following instillation of topical corneal anesthesia (Benox 0.5% with fluorescein dye), IOP measurements were made using the Goldmann applanation tonometer. Three Goldmann applanation tonometer measurement were taken; a minimum 2-min interval was left between IOP measurements to minimize the tonographic effects of repeated tonometry measurements [9]. The mean IOP reading was calculated for each participant and was used in the analyses.

Data analysis

On the basis of previous pilot data from an unrelated data set, it was calculated statistically that a sample of 194 eyes was required to achieve a correlation between CH and age with 80% power at the t-test and at P value less than 0.05 level. In this study, the response variable of interest was either CH or CRF and the predictor variable of interest was only one variable – namely, age. On this basis, 195 participants were randomly selected from the sample of the Egyptian population attending the institute from all governorates, as the institute is a referral hospital.

Ethical considerations

All ethical considerations were respected, including subject information and approval of this noninvasive procedure.


  Results Top


Data were collected from 195 participants and the demographic data are presented in [Table 1]. All participants were Egyptians. Their ages ranged from 19 to 71 years, with an average age of 45 years. There were 44 male and 54 female patients. A total of 98 right eyes and 97 left eyes were tested [Graph 1].
Table 1: Demographic patient data

Click here to view



A high correlation and a low difference between the right and the left eye of the same patient show the accuracy of measurements (t-test, P < 0.001).

The mean CH value was 10.25 ± 0.12 mmHg (range, 6.5–14.4), and the mean CRF was 10.25 ± 0.15 mmHg (range, 4.9–14.2) [Graph 2] and [Table 2]).
Table 2: Values of CH and CRF in normal egyptien

Click here to view



The CH value was lower in older eyes, and the difference between the youngest age group (19–40 years) and the oldest age group (40–71 years) was statistically significant (t-test, P = 0.01).

The mean CH in the youngest age group was 11.1 ± 0.14, and that in the oldest age group was 9.8 ± 0.21 ([Graph 3] and [Table 3]). The mean CRF in the youngest age group was 10.9 ± 0.18, and that in the oldest age group was 10.1 ± 0.19 [Graph 3].
Table 3: Values of CH and CRF in youngest age group and oldest age group of normal egyptien

Click here to view



No significant correlation was found between age and IOP (corneal corrected) (t-test, P = 0.82), or age and IOP (goldman) (t-test, P = 0.11).


  Discussion Top


It has been suggested that CH represents the viscoelastic capacity of the cornea – that is, the cornea's ability to dampen and dissipate applied energy. Hysteresis is dependent on the relative contributions of both elasticity and viscosity, and it has been shown that alterations in either component will have very different and sometimes opposing effects on measured hysteresis [4]. Our finding that CH reduces with age corroborates the findings of experimental ex-vivo studies that show an increase in collagen cross-linking with age [17], which results in a reduction in the viscosity of the cornea and thus an increase in the 'stiffness' of the structure [16]. The data are also in agreement with previous clinical work evaluating the effect of age on ORA-measured corneal biomechanical properties [18]. Taken together, these findings suggest that aging results in an overall reduction in the dampening capacity of the cornea.

CRF was intended to quantify the overall corneal viscoelastic resistance to indentation. The fact that CRF reduces with age is counterintuitive, as it might be expected that the increase in corneal 'stiffness' resulting from an age-related increase in corneal collagen cross-linking would result in an increased resistance to deformation; however, our results do agree with previous findings [19]. The CRF represents a metric of corneal resistance to a near instantaneous indentation force applied axially. Our data suggest that the assumption that CRF reflects overall corneal rigidity may be an oversimplification and that other factors need to be considered when interpreting its value.

One may postulate that CRF represents overall corneal rigidity, as a reduced corneal rigidity would be associated with a reduced scleral rigidity, and so CRF may give an idea about scleral rigidity. However, further work is required to establish the significance of the relationship and how corneal biomechanics relate, if at all, to scleral biomechanics.


  Conclusion Top


In conclusion, this study on corneal biomechanics in normal Egyptian eyes describes the interactions between age and ORA metrics and finds that age is significantly associated with CH and CRF. The mean CH value was 10.25 mmHg (range, 6.5–14.4), and the mean CRF was 10.25 mmHg (range, 4.9–14.2). The mean CH decreases with age, and the mean CRF also decreases with age. However, the variation in both these ORA metrics implies that not only age but there are also other elements contributing to CH and CRF measurement.

Finally, condensing corneal biomechanical measures to a single summary metric will never completely describe the cornea's properties. Recently, investigators evaluating the ORA applanation signal have found that variations in specific signal elements are better descriptors of the corneal response to indentation, particularly following refractive surgery procedures [20]. Further work is required to establish the validity of these new parameters and how they relate to more conventional biomechanical measures.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.

 
  References Top

1.
Schweitzer C, Roberts CJ, Mahmoud AM, Colin J, Maurice-Tison S, Kerautret J. Screening of forme fruste keratoconus with the ocular response analyzer. Invest Ophthalmol Vis Sci 2010; 51:2403–2410.  Back to cited text no. 1
[PUBMED]    
2.
Kotecha A. What biomechanical properties of the cornea are relevant for the clinician? Surv Ophthalmol. 2007; 52(Suppl 2):S109–S114.  Back to cited text no. 2
[PUBMED]    
3.
del Buey MA, Cristóbal JA, Ascaso FJ, Lavilla L, Lanchares E. Biomechanical properties of the cornea in Fuchs' corneal dystrophy. Invest Ophthalmol Vis Sci 2009; 50:3199–3202.  Back to cited text no. 3
    
4.
Lau W, Pye D. A clinical description of Ocular Response Analyser measurements. Invest Ophthalmol Vis Sci 2011; 14:2911–2916.  Back to cited text no. 4
    
5.
Wells AP, Garway-Heath DT, Poostchi A, Wong T, Chan KC, Sachdev N. Corneal biomechanics and optic nerve head compliance: corneal hysteresis but not corneal thickness correlates with optic nerve surface deformation in glaucoma patients. Invest Ophthalmol Vis Sci 2008; 14:3262–3268.  Back to cited text no. 5
    
6.
Dupps WJ Jr. Hysteresis: new mechanospeak for the ophthalmologist. J Cataract Refract Surg 2007; 33:1499–1501.  Back to cited text no. 6
    
7.
Kotecha A, Elsheikh A, Roberts CR, Zhu H, Garway-Heath DF. Corneal thickness- and age-related biomechanical properties of the cornea measured with the ocular response analyzer. Invest Ophthalmol Vis Sci 2006; 47:5337–5347.  Back to cited text no. 7
    
8.
Elsheikh A, Wang D, Brown M, Rama P, Campanelli M, Pye D. Assessment of corneal biomechanical properties and their variation with age. Curr Eye Res 2007; 32:11–19.  Back to cited text no. 8
    
9.
Recep OF, Hasiripi H, Vayisoglu E, Kalayci D, Sarikatipoglu H. Accurate time interval in repeated tonometry. Acta Ophthalmol Scand 1998; 76:603–605.  Back to cited text no. 9
    
10.
Boehm AG, Weber A, Pillunat LE, Koch R, Spoerl E. Dynamic contour tonometry in comparison to intracameral IOP measurements. Invest Ophthalmol Vis Sci 2008; 49:2472–2477.  Back to cited text no. 10
    
11.
Leite MT, Alencar LM, Gore C, Weinreb RN, Sample PA, Zangwill LM, Medeiros FA. Comparison of corneal biomechanical properties between healthy blacks and whites using the Ocular Response Analyzer. Am J Ophthalmol 2010; 150:163–168.  Back to cited text no. 11
    
12.
Glass DH, Roberts CJ, Litsky AS, Weber PA. A viscoelastic biomechanical model of the cornea describing the effect of viscosity and elasticity on hysteresis. Invest Ophthalmol Vis Sci 2008; 49:3919–3926.  Back to cited text no. 12
    
13.
Dupps WJ Jr, Wilson SE. Biomechanics and wound healing in the cornea. Exp Eye Res 2006; 83:709–720.  Back to cited text no. 13
    
14.
Knox Cartwright NE, Tyrer JR, Marshall J. Age-related differences in the elasticity of the human cornea. Invest Ophthalmol Vis Sci 2011; 52:4324–4329.  Back to cited text no. 14
    
15.
Daxer A, Misof K, Grabner B, Ettl A, Fratzl P. Collagen fibrils in the human corneal stroma: structure and aging. Invest Ophthalmol Vis Sci 1998; 39:644–648.  Back to cited text no. 15
    
16.
Elsheikh A, Wang D, Rama P, Campanelli M, Garway-Heath D. Experimental assessment of human corneal hysteresis. Curr Eye Res 2008; 33:205–213.  Back to cited text no. 16
    
17.
Narayanaswamy A, Chung RS, Wu RY, Park J, Wong WL, Saw SM, et al. Determinants of corneal biomechanical properties in an adult Chinese population. Ophthalmology. 2011; 118:1253–1259.  Back to cited text no. 17
    
18.
Mikielewicz M, Kotliar K, Barraquer RI, Michael R. Air-pulse corneal applanation signal curve parameters for the characterisation of keratoconus. Br J Ophthalmol 2011; 95:793–798.  Back to cited text no. 18
    
19.
Kerautret J, Colin J, Touboul D, Roberts C. Biomechanical characteristics of the ectatic cornea. J Cataract Refract Surg 2008; 34:510–513.  Back to cited text no. 19
    
20.
Roberts C. Biomechanics of the cornea and wavefront-guided laser refractive surgery. J Refract Surg 2002; 18:S589–S592.  Back to cited text no. 20
    


    Figures

  [Figure 1]
 
 
    Tables

  [Table 1], [Table 2], [Table 3]



 

Top
 
 
  Search
 
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
Access Statistics
Email Alert *
Add to My List *
* Registration required (free)

 
  In this article
Abstract
Introduction
Aim of the Work
Materials and Me...
Results
Discussion
Conclusion
References
Article Figures
Article Tables

 Article Access Statistics
    Viewed2306    
    Printed193    
    Emailed0    
    PDF Downloaded168    
    Comments [Add]    

Recommend this journal


[TAG2]
[TAG3]
[TAG4]